15-7
Default alarm limits:
Parameter Patient type Upper limit Lower limit
Adult 50 15
Pediatric 50 20
CO2
Neonate 45 30
Adult 4 /
Pediatric 4 /
INS
Neonate 4 /
Adult 30 8
Pediatric 30 8
AwRR
Neonate 100 30
NOTE
z The apnea alarm cannot be disabled.
z When the CO2 monitoring is not required, it is recommended the water trap not
be connected and the work mode be set to STANDBY.
Other Setup
Selecting OTHER SETUP option opens the following menu.
Figure 15-4 CO
2
Other Setup Menu