C-13
C.2.14 Recorder Module Alarm Messages
Alarm message A B L Cause Measure
RECORDER
INIT ERR N
Yes No 2 An error occurs during the
recorder initialization.
Contact the hospital’s
engineers or our customer
Service.
Note: N represents the error number.
RECORDER
SELFTEST
ERR
Yes No 2 An error might occur to
the RAM, ROM and CPU
watchdog.
Open the RECORD menu
and select the CLEAR REC
TASK option. If the problem
remains, contact our
company for repair.
RECORDER
VLT HIGH
No No 1
RECORDER
VLT LOW
No No 1
A problem occurs to the
system power.
If this alarm message is
given for many times,
contact our company for
repair.
RECORDER
HEAD HOT
No No 1 The thermal head of the
recorder is too hot.
Resume the recording till the
recorder cools down
completely. If the problem
still exists, contact our
company for repair.
REC HEAD IN
WRONG POS.
Yes Yes 3 The thermal head of the
recorder is in wrong
position.
Restore the control lever of
the recorder to its previous
position.
RECORDER
OUT OF
PAPER
Yes Yes 3 The recorder paper is used
up.
Replace with a new paper
roll.
RECORDER
PAPER JAM
No No 2 The recording continues
for more than 30 minutes.
Place the recorder correctly
and try again.
RECORDER
COMM ERR
Yes No 2 Error in recorder
communication.
Open the RECORD menu
and select the CLEAR REC
TASK option. If the problem
remains, contact our
company for repair.
TOO MANY
REC TASKS
No No 2 Quite a few alarm events
occur at the same time.
Check the patient’s condition
and the alarms. Open the
RECORD menu and select
the CLEAR REC TASK
option. If the problem
remains, contact our
company for repair.
RECORDER
PAPER W.P.
Yes Yes 2 The paper roll of the
recorder is not placed in
the correct position.
Place the paper roll correctly.
RECORDER S.
COMM ERR
Yes No 2 Error in recorder
communication.
REC NOT
AVAILABLE
No No 2 Error in the recorder work
mode.
Open the RECORD menu
and select the CLEAR REC
TASK option. If the problem
remains, contact our
company for repair.

Table of Contents