4-9
You can perform the following settings in the menu above:
ALM SEL
Alarm selection
Options: COMMON ALM SETUP, XX ALM SETUP;
(XX refers to physiological parameters such as HR,
SPO2, etc.).
If a parameter alarm setup is selected the corresponding
alarm setup items will be displayed in the ALARM
SETUP menu
ALARM VOL
Alarm volume
The volume can be set between 0 and 10. 0 means off
and 10 is the maximum volume.
The volume can not be set to 0 unless the ALARM
SOUND OFF is set to ENABLE (Refer to 4.7
Maintenance). When the ALARM VOL is set to 0, it
will not be saved upon power failure. After the monitor
restarts, the ALARM VOL restores the default. It will
also restore its default if you silence or pause alarms.
REMINDER VOL Options: HIGH, MED, LOW.
ALM REC TIME
Alarm recording time
Options: 8S,16S and 32S.
When an alarm occurs, the recorder records according
to the alarm recording time.
ALM PAUSED TIME Options: 1MIN, 2MIN and 3MIN.