22
Procedures
Press the key.
The load display will become “2000”, and the
digit of minimum display will flash on or off.
When the calibration has completed, the set
value of minimum scale which has registered at
that time will be displayed.
5
By the setting of minimum scale, the digit
of minimum display that flashes on and off
will be as follows:
The minimum scale 1,2,5 10
0
00
0
digit
The minimum scale 10 10
1
11
1
digit
Set the maximum display value with the right
keys.
Setting range for the maximum display will be
(The minimum scale × 100)~99 990.
In order to make good use of the performance,
set within the following range:
: Set value carry key
: Set value digit down key
: Set value Increment key.
: Set value Decrement key.
: Set value Initialization key.
Setting range for maximum display value
100~10 000
200~20 000
500~50 000
1 000~99 990
Minimum Scale
1
2
5
10
By pressing the key continuously,
increment/decrement can be provided
continuously.
6
Press the key.
Load display will show “SMV”.