80
・ FUNC−21 Selection of output logic for BCD data.
(Effective when option is installed.)
Initial value = 000000 000000 = Negative logic
000001 = Positive logic
・ FUNC−22 Selection of output logic for BCD Polarity(POL)
(Effective when option is installed.)
Initial value = 000000 000000 = Negative logic
000001 = Positive logic
・ FUNC−23 Selection of output logic(ERROR, OVR) for BCD flag
(Effective when option is installed.)
Initial value = 000000 000000 = Negative logic
000001 = Positive logic
・ FUNC−24 Operational condition for comparator hysteresis
Initial value = 000001 000000 = On delay
000001 = Off delay
・ FUNC−25 Selection of output logic for BCD P.C.
(Effective when option is installed)
Initial value = 000000 000000 = Negative logic
000001 = Positive logic
・ FUNC−26 Selection of BCD P.C. width.(Effective when option is installed)
Initial value = 000000 000000 = 125 ms approximately.
000001 = 25 ms approximately.
※ Effective only when 4 times of display times/s is selected with FUNC−03. fixed as
approx. 25 ms when 20 time/s of display is selected.
・ FUNC−27 Selection of operational mode for serial interface
(Effective when option is installed.)
Initial value = 000000 000000 = Stream
000001 = Stream + automatic printing
000002 = Stream + manual printing
・ FUNC−28 Selection of output target at the time of RS−232C stream mode.
(Effective when option is installed)
Initial value = 000000 000000 = Display
000001 = Net weight
000002 = Gross weight
※ When “Display” is selected, RS−232C data output interlocks with display.
When “Net weight” is selected, RS−232C data output interlocks with “Net weight”
value at the time of Tare is executed, and also interlocks with “Gross weight” value
at the time of Tare is cancelled.
When “Gross weight” is selected, RS−232C data output interlocks with “Gross
weight” value, whichever Tare is executed or not.
・ FUNC−29 Selection of Operational mode for RS−232C
(Effective when option is installed.)
Initial value = 000000 000000 = Stream
000001 = Command