・ FUNC−50 Selection of target for HOLD
Initial value = 000007 000000 = Ineffective in HOLD
000001 = Holds display only.
000002 = Holds only comparator judgement
display and contact output.
000003 = Holds with the both conditions of
000001 and 000002
000004 = Holds optional output only.
000005 = Holds with the both conditions of
000001 and 000004
000006 = Holds with the both conditions of
000002 and 000004
000007 = Holds with all the conditions of
000001, 000002 and 000004
・ FUNC−51 Lighting up/Putting off bar meter display
Initial value = 000000 000000 = Lighting up
000001 = Putting off
・ FUNC−54 Selection “+display mode” or “+/− display mode”
Initial value = 000000 000000 = + display mode
000001 = +/− display mode
・ FUNC−55 Changeover bridge power supply voltage
Initial value = 000000 000000 = 10 V
000001 = 5 V
000002 = 2.5 V
・ FUNC−56 ON/OFF for short at the input of strain gage applied transducer.
Initial value = 000000 000000 = ON
000001 = OFF
※ Between B and D in strain gage applied transducer can be shorted by the setting
“000001” at the internal of the instrument.
・ FUNC−62 Setting the target for zero tracking
Initial value = 000000 000000 = Gross weight
000001 = Net weight
・ FUNC−97 Setting the data width for stabilized filter
Initial value = 00100 000000 = Stabilized filter OFF
Setting range : 000000 ~ 000999
Unit : 1 D
999 D of data width at “000999” setting
※ Effective only when setting is made between 000001 and 000007 with the FUNC−09,
and also setting is made between 000001 and 00099 with the FUNC−10.