Chapter 5
94
Figure 5-3 Displaying the minimum and maximum thicknesses
The display update rate is independent of the Minimum capture rate. The user can
choose to update the measurements at 4 Hz, 8 Hz, 16 Hz, or 20 Hz. The measurement
capture rate is 60 Hz, and all MIN and MAX thickness values are captured using the
60 Hz capture rate.
The minimum and maximum thickness modes respectively display the smallest and
largest thickness measured value from the time you activate the minimum mode or
from the time you reset it. The modes are useful when it is important to determine the
thinnest/thickest reading obtained while making a series of readings on a test piece.
To activate the minimum, maximum, or min/max mode
1. In the measurement screen, press [MIN/MAX].
2. In the MIN/MAX screen (see Figure 5-4 on page 95):
a) Set ENABLE MIN to ON or OFF.
b) Set ENABLE MAX to ON or OFF.
3. Press [MEAS] to return to the measurement screen.
4. In the measurement screen, press [MEAS] again to reset the held minimum,
maximum, or min/max values.
The thickness display will blank indicating that the old MIN/MAX value is reset.
Saving or sending a MIN/MAX reading also resets the value.
Live/actual thickness
shown in large font
Maximum thickness value
Minimum thickness value