DMTA-10026-01EN, Rev. D, January 2015
Chapter 4
76
86TBM3
(U8771039)
Magnetic target steel ball,
4.76 mm (3/16 in.)
3% 1%
86TBM4
(U8771040)
Magnetic target steel ball,
6.35 mm (1/4 in.)
3% 1%
80TD1
(U8771034)
Flat disk,
12.70 mm (0.500 in.)
3% 2%
80TD2
(U8771035)
Knife edge (V) disk,
6.35 mm (0.250 in.)
3% 2%
86TW1
(U8771041)
Wire target,
1.14 mm (0.045 in.)
3% 2%
86TW2
(U8779858)
Wire target,
0.66 mm (0.026 in.)
3% 2%
Tolerance for metric values: ± [(accuracy × thickness) + 0.003 mm]
Tolerance for English values: ± [(accuracy × thickness) + 0.0001 in.]
Table 9 Calibration accuracy for the 86PR-3 probe
Target
name
Description
Basic
calibration
accuracy
Multipoint
calibration
accuracy
80TB1
(U8771030)
Target steel ball,
1.59 mm (1/16 in.)
4% 3%
80TB2
(U8771031)
Target steel ball,
3.18 mm (1/8 in.)
4% 2%
86TW2
(U8779858)
Wire target,
0.66 mm (0.026 in.)
3% 2%
Tolerance for metric values: ± [(accuracy × thickness) + 0.025 mm]
Tolerance for English values: ± [(accuracy × thickness) + 0.001 in.]
Table 8 Calibration accuracy for 86PR-1 and 86PR-2 probes (continued)
Target
name
Description
Basic
calibration
accuracy
Multipoint
calibration
accuracy