Figure 4-10 Adding additional calibration points
8. Up to eight additional calibration points will be listed on the left side of the
display (see Figure 4-11 on page 82) as follows:
a) Place a specific calibration shim and target onto the probe tip.
b) Press [CAL] to transfer the measurement to the text box.
c) Use the arrows keys to enter the known thickness of the calibration shim, and
then press [CAL].
d) Press [CAL] again to enter the next calibration point.
OR
Press [MEAS] to complete the calibration process.
Figure 4-11 The MULTIPOINT CALIBRATION screen