EasyManua.ls Logo

Olympus Magna-Mike 8600 - Measurements; Figure 4-12 Recalling a Calibration File

Olympus Magna-Mike 8600
186 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA-10026-01EN, Rev. D, January 2015
Chapter 4
84
2. Use the up and down arrow keys to highlight the calibration file to open, and then
press [ENTER].
3. Highlight RECALL, and then press [ENTER] to recall the calibration file. Perform
a Q-CAL by removing the target from the probe tip, and then press [Q-CAL]. The
Magna-Mike 8600 is now ready to make measurements using the target
corresponding to the calibration file that was recalled.
OR
Highlight CANCEL, and then press [ENTER] to cancel the recall of the calibration
file.
The user should verify that the instrument is making accurate thickness
measurements after a calibration file is recalled. Recalled calibration files may have
been performed when the instrument was in a different location, or there may have
been changes in the ambient magnetic field or temperature since the calibration file
was saved.
Figure 4-12 Recalling a calibration file
4.8 Measurements
Thickness measurements can be made once the Magna-Mike 8600 is calibrated for a
particular target: ball, disk, or wire. To measure a test piece, simply place the probe tip
on one side of the material, and place the target on the other side near the probe tip.
IMPORTANT

Table of Contents

Related product manuals