CAUTION
Endodontic treatment must only be performed using the Morita TORX
micromotor's a1 - a3 presets and the Morita CA-10RC-ENDO 10:1
handpiece.
CAUTION
Always check the preset settings before using the micromotor.
The table below lists the factory presets for the Morita TORX micromotor.
These presets can be modified, see section "Modifying preset values" on
page 184.
The tolerance for RPM and torque values is +/- 5%.
NOTE
The torque limit value depends on the handpiece type.
Factory presets for Morita TORX micromotor
Setting Preset 1 - 3 Preset a1 Preset a2 Preset a3 Preset t1 - t3
Instrument
spray
Spray 1:
Water: 50
Air: 80
Spray 2 (Dry):
Water: 0
Air: 80
Default: Spray
off
Disabled Disabled Disabled Spray 1:
Water: 50
Air: 80
Spray 2 (Dry):
Water: 0
Air: 80
Default: Spray
off
Sterile water Available N/A N/A N/A Available
Cooling air Enabled Disabled Disabled Disabled Enabled
Automatic chip
blow
On/Off
Water: 0
Air: 60
Default: Chip
blow on
Disabled Disabled Disabled On/Off
Water: 0
Air: 60
Default: Chip
blow on
Manual chip
blow
Can be
configured to
be used from
foot control
Default: On
N/A N/A N/A Can be
configured to
be used from
foot control
Default: On
Rotation Forward/
Reverse
Default:
Forward
rotation
Forward/
Reverse
Default:
Forward
rotation
Forward/
Reverse
Default:
Forward
rotation
Forward/
Reverse
Default:
Forward
rotation
Forward/
Reverse
Default:
Forward
rotation
23 Operating instruments