unit are flushed with water the next morning. The disinfectant should not be
left in the waterlines for longer than one night.
After installation, or if the dental unit has not been used for a long time, the
waterlines of the dental unit must be treated with a disinfectant for three
consecutive cycles.
If you have several dental units with Waterline Cleaning System (WCS), you
may use the same container in all of them.
WARNING
Even though every effort has been taken to ensure patient safety, even in
case of malfunction or misuse, always make sure that the unit is properly
flushed before taking it into use.
CAUTION
Only Planmeca Planosil, Planmeca PlanPure or Alpro Bilpron disinfectant
must be used. Planmeca does not guarantee the suitability of and is not
liable for damages caused by other disinfectants.
NOTE
If you are using the Clean Water System (CWS), clean the waterlines
according to the instructions in section "Waterline cleaning" on page 241.
NOTE
The dental unit waterlines should be cleaned once a week; the Maintenance
main window displays when cleaning programs were started. If Planmeca
Romexis Clinic Management software is used, detailed information on
cleaning cycles can be retrieved. For more information, see the Planmeca
Romexis user’s manual.
NOTE
Do not let the Waterline Cleaning System run over the weekend.
NOTE
The water must flow properly through the cup fill line, otherwise the unit will
not be flushed. Make sure that the cup fill line is not totally closed. The flow
through the cup fill line is recommended to be approximately 1 dl / 5 sec.
27.6.2 Starting waterline cleaning cycle
About this task
NOTE
The waterline cleaning program can be cancelled by pressing Cancel. If the
cleaning cycle is cancelled after the disinfectant container has been attached
to the dental unit, flushing is required. Help message H 35 guides you
through the flushing.
NOTE
The pop-up windows can be minimised by pressing Minimise.
27 Flushing and cleaning programs
User's manual Planmeca Compact i 215

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact i Touch v2 and is the answer not in the manual?

Planmeca Compact i Touch v2 Specifications

General IconGeneral
BrandPlanmeca
ModelCompact i Touch v2
CategoryMedical Equipment
LanguageEnglish

Related product manuals