31.3.6 Suction holder related error messages
Suction holder related error messages
ERROR
CODE
EXPLANATION
E47 Suction holder is of new type.
E47.1 Suction holder in branch 1 is of new type.
Update MCB software.
E47.2 Suction holder in branch 2 is of new type.
Update MCB software.
E48.1 No suction holders detected.
Either the holder is missing or faulty, or the cable to the holder
is loose or faulty.
E48.2 Suction holder related error messages.
E48.3 Suction holder related error messages.
E48.4 Suction holder related error messages.
E49.1 Suction holder PCB or its cable is faulty in branch 1.
Either the holder is missing or faulty, or the cable to the holder
is loose or faulty.
E49.2 Suction holder PCB or its cable is faulty in branch 2.
Either the holder is missing or faulty, or the cable to the holder
is loose or faulty.
E50 Unidentified STCS jumper configuration.
E50.1 Water management system (WMS) must be set to Bottle
water off -mode when starting suction cleaning.
E50.2 Suction disinfectant pump is not running.
31.3.7 Foot control related error messages
Foot control related error messages
ERROR
CODE
EXPLANATION
E51.1 Cable from foot control is loose or connected into wrong
connector on MCB.
E51.2 Foot control is not responding.
Ensure that the foot control cable is properly attached to the
unit. If it is, contact service.
E51.3 Foot control pedal was pressed down when dental unit was
switched on.
Restart the unit without touching the foot control.
E51.4 Foot control pedal was not in home position when dental unit
was switched on.
Restart the unit without touching the foot control.
31 Help and error messages