Instrument related error messages
ERROR
CODE
EXPLANATION
E23.6 Instrument overvoltage error.
E23.7 Instrument data communication error.
E23.8 Apex locator / Morita micromotor PCB failure.
E23.9 Unable to save instrument settings to selected preset.
E25 Satelec mini LED polymerisation light requires newer IMUX
PCB version.
31.3.5 Instrument multiplexer related error messages
Instrument multiplexer related error messages
ERROR
CODE
EXPLANATION
E28.1 Reference resistor signal out of bounds when unit is switched
on.
E28.2 Active instrument signal out of bounds.
Error in hose identifcation system in IMUX. Ensure that the
instrument's quick connector is properly attached.
E28.3 Syringe signal out of bounds.
Error in hose identifcation system in IMUX. Ensure that the
instrument's quick connector is properly attached.
E28.4 Reference resistor signal out of bounds when instrument is
activated.
Ensure that the instrument's quick connector is properly
attached.
E28.5 Syringe signal changes during use (blinking).
Error in hose identifcation system in IMUX. Ensure that the
instrument's quick connector is properly attached.
E29 Voltage error in instrument light power supply in IMUX.
If there is a problem with one instrument only, replace the
hose. If the problem is with several instruments or the hose is
ok, contact service.
E32.1 IMUX is not responding. IMUX cable disconnected or failure in
cable or IMUX.
E32.2 IMUX software is not compatible (IMUX is of new type) with
MCB software.
Update MCB software.
E32.3 IMUX data error (cable short circuit).
E32.4 IMUX communication error.
E32.5 Wrong IMUX type or wrong unit type.
E33.1 Syringe select valve short circuit.
E33.2 Instrument 1 select valve short circuit.
E33.3 Instrument 2 select valve short circuit.
31 Help and error messages
User's manual Planmeca Compact i 269