The system will first identify the instruments and then each instrument
hose is flushed with air and water in its turn. The time of flushing is the
same for all instruments. The total flushing time is displayed on the
control panel. At the same time the cup fill tube and the bowl are flushed,
too.
NOTE
Adjust the water flow of the cup fill tube so that the water does not splash
into the bowl.
6. Release the instrument arms (if your dental unit is equipped with
balanced instrument arms).
7. Once the flushing cycle is finished, H 36 is displayed. Remove the
instruments from the flushing holder and return them to the instrument
console. Place the flushing holder cover back to its position.
If water flow was not detected for all water consuming instruments,
H 36.1 is displayed. Run instruments for a while with spray water to
ensure that the flushing of instruments is sufficient.
What to do next
NOTE
After the flushing cycle, the unit should immediately be switched off. This
procedure ensures that cool water remains in the pipelines of the unit, thus
minimising the growth of biofilm.
NOTE
The dental unit can be configured to display help message H 99 if the unit
has been switched off without performing long flushing after water has been
used. When H 99 has been enabled, it can be displayed as a reminder only,
or it can be set to disappear only after you have performed long flushing. To
take help message H 99 into use, contact your Planmeca dealer.
27.5 Suction cleaning
About this task
NOTE
The Suction Tube Cleaning System (STCS) is an optional feature. If your
dental unit does not have this feature, clean the suction tubes with OroCup.
For more information, see section "OroCup" on page 235.
Steps
1. Remove the suction handpieces from the suction tubes and clean them
according to the manufacturer’s instructions.
2. Press Maintenance to go into maintenance mode.
A pop-up window opens.
3. Select Suction cleaning.
27 Flushing and cleaning programs
212 Planmeca Compact i User's manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact i Touch v2 and is the answer not in the manual?

Planmeca Compact i Touch v2 Specifications

General IconGeneral
BrandPlanmeca
ModelCompact i Touch v2
CategoryMedical Equipment
LanguageEnglish

Related product manuals