EasyManua.ls Logo

Planmeca Compact i Touch v2 - Duration of Bowl Rinsing

Planmeca Compact i Touch v2
314 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1. Adjustment knob for bowl flush
2. Adjustment knob for cup fill
When you have adjusted the flow rates, you might need to adjust the
duration of the bowl rinse and cup fill. For instructions, see sections
"Duration of bowl rinsing" on page 188 and "Duration of cup filling" on page
189.
25.6.2 Duration of bowl rinsing
Steps
1. Press Program.
2. Press Bowl rinse.
A pop-up window opens.
3. Adjust the duration of the bowl rinsing.
The minimum value is 5 seconds and the maximum 240 seconds, and it
is adjusted in steps of 5.
4. Press OK.
The pop-up window closes.
25 Programming
188 Planmeca Compact i User's manual

Table of Contents

Related product manuals