28.5.3 After each working day
Steps
1. Disinfect the suction system either by running the suction cleaning
program or by using OroCup. For information on the suction cleaning
program, see section "Suction cleaning" on page 212.
2. When the suction system has been rinsed, disinfect the suction
handpieces in a washer-disinfector at 93°C, then optionally autoclave
them at 134°C.
3. Flexy-holder: Disinfect the supplementary holders, suction tube holders
(including rollers) and instrument holder in a washer-disinfector at 93°C.
4. Flexy-holder: Wipe the holder for the intraoral scanner with Planmeca
approved surface disinfectant.
28.5.3.1 OroCup
About this task
The following describes how to disinfect the suction system with OroCup.
Steps
1. Mix the cleaning solution:
Pour 20 ml Planmeca approved suction disinfectant into the rinsing
bottle. Add 1 l water and shake well.
NOTE
Do not use dish washing detergents.
28 Cleaning and disinfection
User's manual Planmeca Compact i 235