Disinfect the flushing holder once a month in a washer-disinfector at 93°C or
autoclave it at 134°C.
Note that the flushing holder cover can not be autoclaved. Remove the cover
before autoclaving the instrument flushing holder.
28.4.3 Suction tube cleaning holder
Disinfect the suction tube cleaning holder once a month in a washer-
disinfector at 93°C.
28.4.4 Clean-water bottle
Use a bottle brush and a mild soap- and water solution to clean the clean-
water bottle once a week.
To remove the clean-water bottle from the dental unit, first turn the Water
bottle switch toward 'OFF'. Then, take a hold of the bottom of the bottle, turn
the bottle slightly so that the pin moves along the groove and pull the bottle
downwards.
28 Cleaning and disinfection
User's manual Planmeca Compact i 233