Factory presets for Morita TORX micromotor
Setting Preset 1 - 3 Preset a1 Preset a2 Preset a3 Preset t1 - t3
Instrument
light
On/Off
Brightness:
100
Default:
Instrument
light on
N/A N/A N/A On/Off
Brightness:
100
Default:
Instrument
light on
Power curve Linear/Full
Default: Linear
Linear/Full
Default: Full
Linear/Full
Default: Full
Linear/Full
Default: Full
Linear/Full
Default: Linear
Rpm Value range:
100 - 40 000
Default: 40 000
Value range:
100 - 1000
Value range:
100 - 1000
Value range:
100 - 1000
Value range:
100 - 2000
Rpm limit Enabled/
Disabled
Value range:
100 - 36 000
rpm
Default:
Disabled
Enabled
Value range:
100 - 1000 rpm
Default: 400
rpm
Enabled
Value range:
100 - 1000 rpm
Default: 400
rpm
Enabled
Value range:
100 - 1000 rpm
Default: 1000
rpm
Enabled
Value range:
100 - 2000 rpm
Default: 2000
rpm
Torque limit Disabled Enabled/
Disabled;
autoforward
Value range:
0.5 - 5.0 Ncm
Default:
Enabled, 1.0
Ncm
Enabled/
Disabled;
autoforward
Value range:
0.5 - 5.0 Ncm
Default:
Enabled, 2.0
Ncm
Enabled/
Disabled;
autoforward
Value range:
0.5 - 5.0 Ncm
Default:
Disabled
Enabled;
autoforward
Value range:
0.5 - 2.5 Ncm
Default:
Enabled, 2.5
Ncm
Handpiece 1:1 10:1 10:1 10:1 1:1
Apical
measurement
N/A Available Available Available N/A
Apical actions N/A Off/ Stop/
AutoForward
Default:
AutoForward
Off/ Stop/
AutoForward
Default:
AutoForward
Off/ Stop/
AutoForward
Default: None
N/A
Apical beep
volume
N/A On/off
Default: On
On/off
Default: On
On/off
Default: On
N/A
23.6 Apex locator
The Morita Root ZX mini U apex locator can be used as a guide when
measuring the root canal. It is available for the Morita TORX micromotor, the
CA-10RC-ENDO 10:1 handpiece and for hand files.
The apex locator can be operated from the dental unit's control panel. More
detailed information on how to measure the root canal can be found in the
operation instructions provided by Morita.
23 Operating instruments
User's manual Planmeca Compact i 145