25.4 Instrument settings
When you have signed in to the dental unit with your own user name, all
changes you make to the instrument settings and then save, are saved to
your personal settings. This means that whenever you sign in to the dental
unit, you will use your last saved instrument settings.
25.4.1 Instrument speed/power limit
About this task
NOTE
The instrument speed/power limit does not affect the air driven instruments
for which quickstart has been selected.
Steps
1. Activate the instrument.
2. Press Program.
3. Press Instrument speed/power limit.
When you are programming the turbine, press Turbine speed/power
limit.
A pop-up window opens.
4. For W&H and Minetto brushless micromotors only: Press %/rpm to
select whether you want to adjust the instrument speed/power limit in
percentage points or in rpm. The appearance of the button in the
instrument view changes accordingly.
5. Adjust the instrument speed/power limit.
The minimum speed/power limit is 5 or 10% (depending on the
instrument) and the maximum is 95%. The adjustment step is 5.
If the preset value is, for example, 50%, the foot control pedal movement
controls the speed/power between 5 - 50% instead of the normal range
of 5 - 100%.
6. Press OK.
The pop-up window closes.
25.4.2 Instrument spray
Steps
1. Activate the instrument.
25 Programming
176 Planmeca Compact i User's manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact i Touch v2 and is the answer not in the manual?

Planmeca Compact i Touch v2 Specifications

General IconGeneral
BrandPlanmeca
ModelCompact i Touch v2
CategoryMedical Equipment
LanguageEnglish

Related product manuals