How to clean dental unit surfaces
When Part Cleaning agent Additional cleaning method
Dish-
washer
(65°C)
Washer-
disinfector
(93°C)
Autoclave
(134°C)
Tray tables Planmeca
approved surface
disinfectant
Flexy-holder,
tablet holder and
suction arm
Planmeca
approved surface
disinfectant
Suction
handpieces
Planmeca
approved surface
disinfectant
X X
Suction tubes Planmeca
approved surface
disinfectant
Monitor Planmeca
approved surface
disinfectant
Upholstery Mild soap- and
water solution
After the working
day
Upholstery Planmeca
approved
upholstery
disinfectant
On Flexy-holder:
Suction tube
holders (incl.
rollers),
instrument holder
and
supplementary
holders
Planmeca
approved surface
disinfectant
X
On Flexy-holder:
Holder for
intraoral scanner
Planmeca
approved surface
disinfectant
Weekly Upholstery Planmeca
approved
upholstery
treating agent.
Wipe away any
excess oil after
treatment.
Instrument console
Before cleaning the instrument console, lock the touch display by pressing
the Lock button in the
Maintenance
window.
28 Cleaning and disinfection
230 Planmeca Compact i User's manual