Help messages
HELP
CODE
MEANING ACTION OR EXPLANATION
H 92 Help (Stop) button pressed.
H 93 This button has no function at the moment
because there is no motorised light.
H 94 Clock calibration failed. Maximum
calibration range exceeded.
H 95 Patient chair automatic position cannot be
reached because legrest is unlocked or
armrests are blocking movement.
H 96 PlanID tag already in use. PlanID tag assigned to another user.
H 96.1 Different user already logged in. The unit has rejected a login request
because a different user is already logged
in.
H 97 Failed to assign PlanID tag to user. Check Romexis connection.
H 98 User not found. User not found with this PlanID tag. Create
a new user or select an existing Romexis
user and assign the PlanID tag.
H 98.1 User login failed. Retry login or reboot the unit. If problem
persists, contact service.
H 98.2 RFID unknown. The RFID is unknown to the system.
H 99 Unit has been switched off without
performing long flushing after water has
been used. Perform long flushing.
This help message can be disabled, contact
your Planmeca dealer.
HE200 Detected new software update. Confirm to
install?
HE201 Detected new software update SD-card.
Confirm to install 5.3.0.9.R ?
HE210 Software update. Installing to Control panel
bootloader sw. DO NOT TURN POWER
OFF.
Do not turn the power off while the software
is being updated.
HE212 Software update. Installing to Control panel
application sw. DO NOT TURN POWER
OFF.
Do not turn the power off while the software
is being updated.
HE214 Software update. Installing to Control panel
application sw. DO NOT TURN POWER
OFF.
Do not turn the power off while the software
is being updated.
HE216 Software update. Downloading from server.
DO NOT TURN POWER OFF.
Do not turn the power off while downloading
software from the server.
HE218 Software update. Installing to Main PCB.
DO NOT TURN POWER OFF.
Do not turn the power off while the software
is being updated.
END Flushing completed. Remove syringe from
flushing holder and return it to console. If
syringe is removed, return it to flushing
holder.
DOOR Cuspidor door is open. Close the door.
31 Help and error messages
User's manual Planmeca Compact i 263