5 Annual maintenance
To guarantee the dental unit’s proper operation, the unit must be checked
and serviced by a qualified Planmeca service technician according to the
maintenance schedule that has been set for your dental unit.
In the annual maintenance, the service technician replaces all parts
specified by the maintenance kit. These include, but are not limited to, parts
in contact with air, water and suction systems. In addition, the service
technician checks and services all dental unit parts suspect to wear and tear
in normal use. These include parts in the junction box, foot control,
instrument console, delivery arms, patient chair, suction arm and operating
light. Also, the mechanical stability and electrical safety inspection is
performed.
The default maintenance interval is 365 days.
A help message will remind you about the annual maintenance well in
advance.
5 Annual maintenance