The tray table is attached to the instrument console with a quick-connector,
enabling an easy attachment and detachment of the tray table.
10.6 Sterile water system
10.6.1 Introduction
When the sterile water system is used, sterilised water is fed from a
disposable sterile water bag through an external disposable tube to a
surgical handpiece or scaler. The instrument must have an externally
mounted spray nozzle on which the sterile water tube is fitted.
CAUTION
Before using an instrument with sterile water, make sure water comes out
from the instrument.
CAUTION
The sterile water system can only be used with surgical handpieces that are
intended for sterile use, as well as with the Satelec Newtron, Satelec
Newtron LED and the LM-ProPower SteriLED scalers. If other instruments
are used, the system is not sterile.
CAUTION
To ensure sterile conditions, extra precaution should be taken to make sure
that correct procedures are followed and all the components and tools used
(e.g. scissors) are sterile.
CAUTION
The sterile water bag, nozzle and tubes are disposable and intended for
single use only.
10 Instrument system
34 Planmeca Compact i3 User's manual