6. Press OK.
The pop-up window closes and the new date is displayed on the control
panel.
24.10 Adjusting maintenance functions
About this task
You can adjust the following maintenance functions in the
Program -
Maintenance
window:
duration of short flushing
duration of long flushing
duration of manual suction cleaning.
Alternatively, the values can be adjusted in the service modes, contact your
Planmeca dealer.
Steps
1. Press Program.
2. Press Maintenance.
The
Program - Maintenance
window opens.
3. Press Edit next to any item to open a new window where you can adjust
the disinfectant concentration and/or duration of maintenance programs.
A pop-up window opens.
4. Adjust the value by dragging the slider handle.
Dragging the slider handle to the right increases the value and dragging
it to the left decreases the value.
NOTE
Adjusting the value of the manual suction cleaning time to smaller than
60 s disables the feature and removes it from the maintenance menu.
Default value Value range
Duration of short flushing 30 s / hose 10 - 30 s / hose
Duration of long flushing 2 min total 2 - 9 min total
Duration of manual
suction cleaning
2 min total 1 - 7 min total
24 Programming
User's manual Planmeca Compact i3 169

Table of Contents

Other manuals for Planmeca Compact i3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact i3 and is the answer not in the manual?

Planmeca Compact i3 Specifications

General IconGeneral
BrandPlanmeca
ModelCompact i3
CategoryDental equipment
LanguageEnglish

Related product manuals