If necessary, the handpiece can be completely disassembled for more
thorough cleaning. The handpiece parts can be disinfected in a washer-
disinfector at 93°C, then optionally autoclaved at 134°C.
Saliva suction handpiece
The saliva suction handpiece can be disassembled for cleaning by pulling
out the end of the handpiece and removing the core (2) and the adjuster (1).
The handpiece parts can be disinfected in a washer-disinfector at 93°C, then
optionally autoclaved at 134°C.
27.9 Planmeca intraoral scanner
For information on how to clean the Planmeca intraoral scanner, see
Planmeca FIT user’s manual
.
27.10 External PC
The external PC together with its mouse and keyboard can be wiped with a
dry cloth, or according to the manufacturer’s instructions.
27 Cleaning and disinfection
User's manual Planmeca Compact i3 191