NOTE
To avoid stains, possible disinfectant splashes should be wiped away
immediately.
28.2.1 After the working day
About this task
NOTE
You can interrupt the flushing cycle by pressing Close. After the interruption,
the dental unit can be used normally.
Steps
1. Remove the clean-water bottle.
First, turn the Water bottle switch toward 'OFF'. Then, take a hold of the
bottom of the bottle, turn the bottle slightly so that the pin moves along
the groove and pull the bottle downwards.
2. Empty the possible water from the clean-water bottle.
3. Fill the clean-water bottle with approx. 2 dl of Planmeca approved
waterline disinfectant. The amount depends on the unit configuration.
The assistant's syringe uses about 1 dl of disinfectant. Thus, if the
syringe is in use, add this amount to your calculations on the required
disinfectant amount.
We recommend you keep record of how much disinfectant has been
used.
28 Clean water system
User's manual Planmeca Compact i3 195