7. Press Program.
8. Press Torque to adjust the torque limit rate.
A pop-up window opens.
9. Adjust the torque limit rate.
The adjustment range is 10% - 95%.
10. Press OK.
The pop-up window closes and the new value is displayed on the control
panel.
24.4.7 Bien-Air MX2 micromotor
24.4.7.1 Modifying preset values
About this task
CAUTION
Before modifying the torque and RPM functions, please refer to the file
manufacturer’s instructions concerning torque and RPM value
recommendations for the files.
If you want to modify the presets presented in section "Presets" on page
115, you can do so by following the steps below.
Steps
1. Activate the micromotor.
2. Press Torque.
3. Select the micromotor’s drive mode.
Toggle between the drive modes by pressing the current drive mode -
button.
The modes are:
Auto stop
The micromotor stops when the torque limit is
reached.
Auto reverse
When the torque limit is reached the micromotor will
operate counter-clockwise.
24 Programming
156 Planmeca Compact i3 User's manual

Table of Contents

Other manuals for Planmeca Compact i3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact i3 and is the answer not in the manual?

Planmeca Compact i3 Specifications

General IconGeneral
BrandPlanmeca
ModelCompact i3
CategoryDental equipment
LanguageEnglish

Related product manuals