2. When the suction system has been rinsed, disinfect the suction
handpieces in a washer-disinfector at 93°C, then optionally autoclave
them at 134°C.
3. Flexy-holder: Disinfect the supplementary holders, suction tube holders
(including rollers) and instrument holder in a washer-disinfector at 93°C.
4. Flexy-holder: Wipe the holder for the intraoral scanner with Planmeca
approved surface disinfectant.
27.8.4 Cleaning suction handpieces
The following explains how the suction handpieces can be disassembled for
cleaning.
NOTE
We recommend that the suction handpieces be replaced once a year.
High-volume suction handpiece
Unscrew the handpiece from the suction tube for cleaning.