General error messages
ERROR
CODE
EXPLANATION
E3.5 Internal air pressure (after pres. regulator) is too high.
Check pressure regulator, pressure sensor and pressure
block connector.
E3.6 Waterline cleaning feed pressure too low.
Check and adjust waterline cleaning feed pressure. Replace
waterline cleaning feed regulator if problem still exists.
E3.7 Waterline cleaning flush pressure too low.
Check dental unit's incoming water pressure. Ensure that
dental unit inlet water tap is open.
E4.1 Short circuit in cable from MCB to control panel / suction
holder in branch 1.
E4.2 Short circuit in cable from MCB to control panel / suction
holder in branch 2.
E4.3 Communication bus (CAN) is not operational or CAN devices
are disconnected (foot control, operation light, motorised
headrest).
E4.4 Communication bus (Console CAN) is not operational.
E5 Unsupported device attached.
Detach the device or contact service.
E6.01 Cuspidor up -safety circuit error.
E6.02 Cuspidor down -safety circuit error.
E6.03 Chair down -safety circuit error.
E6.04 Bowl up -safety circuit error.
E6.05 Bowl down -safety circuit error.
E6.07 Console up -safety circuit error.
E6.08 Console down -safety circuit error.
E6.09 Emergency stop -safety circuit error.
29.3.2 Power supply related error messages
WARNING
If a fuse has blown, contact your Planmeca dealer. Fuses must be changed
only by a qualified Planmeca service technician.
Power supply related error messages
ERROR
CODE
EXPLANATION
E7 Power supply related error messages. IPS voltage failure on
MCB.
E9 Internal operating voltage too low. Unit functionality may be
limited.
29 Help and error messages
User's manual Planmeca Compact i3 213