General error messages
ERROR
CODE
EXPLANATION
E1.11 Separator start signal / Microvac's start valve short circuit.
E1.12 Assistant syringe water valve short circuit.
Disconnect valve and check if error disappears.
E1.13 Separator alarm reset signal short circuit.
Disconnect output and check if error disappears.
E1.14 PLANET output (ELMP) short circuit.
Disconnect output and check if error disappears.
E2.01 Incoming water valve open circuit.
Error can be detected when circuit is dead.
E2.02 Incoming air valve open circuit.
Error can be detected when circuit is dead.
E2.03 Bowl rinse valve open circuit.
Error can be detected when circuit is dead.
E2.04 Cup filling valve open circuit.
Error can be detected when circuit is dead.
E2.05 Main Control PCB output open circuit warning.
E2.06 Separator pulse valve open circuit.
Error can be detected when circuit is dead.
E2.07 Main Control PCB output open circuit warning.
E2.08 Main Control PCB output open circuit warning.
E2.09 Main Control PCB output open circuit warning.
E2.10 Main Control PCB output open circuit warning.
E2.11 Separator start signal open circuit / Microvac valve open
circuit.
E2.12 Assistant syringe water valve open circuit.
Error can be detected when circuit is dead.
E2.13 Separator alarm reset signal open circuit.
Error can be detected when circuit is dead.
E2.14 Main Control PCB output open circuit warning.
E2.15 Not used or not in the manuals.
E3.1 Incoming air pressure compared to water pressure is too low.
Ensure that air pressure main switch is on. If problem persists,
contact service.
E3.2 Internal water pressure too low.
Check dental unit's incoming water pressure. Ensure that
dental unit inlet water tap is open.
E3.4 Internal air pressure (after regulator) is too low.
Ensure that air pressure main switch is on. If problem persists,
contact service.
29 Help and error messages
212 Planmeca Compact i3 User's manual

Table of Contents