EasyManuals Logo

Planmeca Compact i3 User Manual

Planmeca Compact i3
252 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #184 background imageLoading...
Page #184 background image
Steps
1. Press Maintenance to go into maintenance mode.
2. Remove the instrument handpieces from the instruments and clean them
according to the manufacturer’s instructions.
3. Place all water consuming instruments (including syringes) into the
openings in the instrument flushing holder.
Optionally, in dental units with balanced instrument arms, you can
remove the hoses of the instruments to be flushed from the instrument
arms / instrument holders before placing the instruments in the flushing
holder.
In dental units with hanging-tube instruments, the instruments to be
flushed must be removed from the instrument holders before placing
them in the flushing holder.
Each type of syringe has its dedicated place in the flushing holder. The
DCI and Luzzani Minibright syringes require an adapter that must be
placed in the instrument flushing holder to keep the syringe in place. The
picture below shows an example of how the syringes can be placed, but
note that the picture does not represent the actual holder.
26 Flushing and cleaning programs
176 Planmeca Compact i3 User's manual

Table of Contents

Other manuals for Planmeca Compact i3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact i3 and is the answer not in the manual?

Planmeca Compact i3 Specifications

General IconGeneral
BrandPlanmeca
ModelCompact i3
CategoryDental equipment
LanguageEnglish

Related product manuals