CAUTION
If the drain is blocked, the dental unit might overflow with contaminated
water and excessive water could flow onto the floor. Shut down the unit and
contact your property's caretaker and your Planmeca dealer.
CAUTION
For surgical procedures, use sterile irrigating solutions, such as sterile water
or saline. Appropriate delivery devices should be used to deliver sterile
irrigating solutions during surgery. This may include a dedicated surgical
irrigation system with components including handpieces that are single-use
disposable or compatible with heat sterilisation methods used in outpatient
dental settings.
CAUTION
A power cut will shut down the software-controlled backflow prevention
system. If you are using a turbine without a built-in backflow prevention
system, contaminated water can enter into the turbine and turbine hose in
the event of a power cut.
CAUTION
In extreme operating conditions the temperature of the chair lift and backrest
motors might rise significantly. DO NOT TOUCH THE MOTORS!
CAUTION
In extreme operating conditions the surface temperature of the seat
upholstery may rise to 44°C (111°F). When applying maximum allowed load
on the patient chair in hot climate conditions, ensure sufficient time for the
seat upholstery to cool down.
NOTE
Before switching on the dental unit, make sure that the main water feed, air
pressure and suction motor are turned on.
NOTE
National regulations concerning the quality of dental water and dental air
must be followed when using the Planmeca dental unit.
NOTE
The water used by the dental unit instruments are to be used for rinsing only.
For more information, please contact your Planmeca dealer.
NOTE
The main water feed must be turned off when the dental unit is not in use.
NOTE
If using air abrasives, please consult instructions for use given by the air
polisher manufacturer. Make sure that adequate flushing of dental unit is
performed immediately after using air polishers.
NOTE
The air used by the dental unit instruments must be dry, clean and oil-free.
7 For your safety
User's manual Planmeca Compact i3 11

Table of Contents

Other manuals for Planmeca Compact i3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact i3 and is the answer not in the manual?

Planmeca Compact i3 Specifications

General IconGeneral
BrandPlanmeca
ModelCompact i3
CategoryDental equipment
LanguageEnglish

Related product manuals