Steps
1. Mix the cleaning solution:
Pour 20 ml Planmeca approved suction disinfectant into the rinsing
bottle. Add 1 l water and shake well.
NOTE
Do not use dish washing detergents.
2. Place the rinsing bottle (for example OroCup) on a flat surface (table or
floor).
3. Remove the suction handpieces from the suction tubes and clean them
according to the manufacturer’s instructions.
4. Press Maintenance to go into maintenance mode.
A pop-up window opens.
5. Remove the suction handpieces from their holders and push them onto
the inserts inside the cap of the rinsing bottle.
6. Press Manual suction cleaning.
7. Return the suction tubes back to the holder after suction cleaning has
stopped.
8. Replace the exchangeable parts (filters etc.) the next morning.
26 Flushing and cleaning programs