Help messages
HELP
CODE
MEANING ACTION OR EXPLANATION
H 52 Patient chair can not be driven upward
because safety switch of side delivery arm
is released.
H 53 Patient chair can not be driven downward
because safety switch of side delivery arm
is released.
H 54 Cup is inserted in cup holder. Remove the cup and make sure the cup fill
tube is above the bowl.
H 55 Disinfectant container attached. To start
waterline cleaning, select instruments and
press Waterline cleaning button.
H 57 Vision action failed. Select patient in
Romexis.
Select patient in Planmeca Romexis before
performing Solanna Vision action.
H 60 WMS: Container almost empty in Domestic
water supply -mode.
Reduce water use for a moment.
H 61 WMS: Container almost empty in Bottled
water -mode.
Reduce water use and fill the container as
soon as possible.
H 62 WMS: Container switch is turned off. Switch on the container switch.
H 63 Cuspidor door is open. Close the door.
H 64 Fill container with disinfectant over upper
sensor.
H 65 Fill container with water over upper sensor.
H 66 Wait 8 hours or switch unit off for the night.
H 67 Instrument flush cancelled. Return
instruments to instrument console.
H 68 Romexis connection lost. Make sure the Ethernet cable is attached
and Romexis is running.
H 69 Unable to create a new user. Maximum
amount of local users exceeded.
H 70 Brushless micromotor is stuck or broken.
H 71 Romexis user ID already in use. Check the ID, select another one, or allow
rebind in Romexis.
H 72 Incompatible Romexis version. PlanID is not supported with this version of
Romexis.
H 72.1 Incompatible Romexis version User settings for currently active instrument
can not be used. Romexis version does not
support user instrument settings or presets
for currently active instrument. Check the
instrument settings before using instrument
and update Romexis.
H 73 Failed to copy user profile from Romexis.
H 73.1 Failed to save user settings.
H 74 PlanID reader missing. PlanID reader is either missing or not
functional.
29 Help and error messages
User's manual Planmeca Compact i3 207

Table of Contents

Other manuals for Planmeca Compact i3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact i3 and is the answer not in the manual?

Planmeca Compact i3 Specifications

General IconGeneral
BrandPlanmeca
ModelCompact i3
CategoryDental equipment
LanguageEnglish

Related product manuals