The holder is attached on the delivery arm.
2. Open the lid of the water pump and place the silicone part of the sterile
water tube in the pump.
Make sure that the sterile water tube is inserted so that water is pumped
from the sterile water bag to the instrument. The arrow on the pump
indicates the direction of the water flow, away from the bag and toward
the instrument.
3. Close the lid of the pump.
4. Connect the sterile water tube to the sterile water bag by pushing the
nozzle firmly into the opening at the bottom of the sterile water bag.
5. Route the other end of the tube to the instrument hose.
10 Instrument system
36 Planmeca Compact i3 User's manual