Chapter 11 -
V
ENTILATOR
C
HECKOUT
T
ESTS
This chapter details five test procedures accessed through the Vent Check menu and used to verify
the proper operation of the LTV
®
. These tests should be performed as shown in the schedule in
Appendix B - Set Up / Maintenance for periodic maintenance and testing of the ventilator.
The five test procedures are:
Test Test used to:
Alarm Test Verify that the audible alarm is working correctly.
Display Test Verify that the ventilator displays are working correctly.
Control Test Verify that the buttons and the Set Value knob are working correctly.
Leak Test Test the patient circuit for leaks.
Vent Inop Alarm Test Verify that the Inop Alarm is working correctly.
The Vent Check Menu is set up as follows:
VENT CHECK
ALARM
DISPLAY
CONTROL
LEAK
EXIT
WARNING
Ventilator Checkout Tests –Gas is not delivered to the patient during these tests. Disconnect the
patient from the ventilator and ventilate using an alternative method before running the Ventilator
Checkout tests.
Leak Testing the Patient Breathing Circuit – The patient circuit must be leak tested in VENT
CHECK mode before connection to the patient. Ventilator Checkout mode should also be used to
check for correct operation of the alarms, displays and controls. Harm to the patient or ineffective
ventilation may result from failure to leak test the patient breathing circuit before connection to a
patient. When using a heated humidifier, include it in the circuit when leak testing.
AVERTISSEMENT
Tests de vérification du ventilateur – Noter que le gaz n’est pas transmis au patient au cours de
ces tests. Débrancher le patient du ventilateur et ventiler à l’aide d’une forme de ventilation
alternative avant de procéder aux tests de vérification du ventilateur.
Contrôle de l’étanchéité du circuit respiratoire du patient – L'étanchéité du circuit respiratoire du
patient (vérification de ventilation) doit être vérifiée en mode VENT CHECK avant le raccordement au
patient. En outre, on doit utiliser le mode Ventilator Checkout (vérification du ventilateur) afin de
s’assurer du fonctionnement adéquat de l’alarme, des affichages et des commandes du ventilateur.
Le défaut de vérifier l’étanchéité du circuit respiratoire du patient avant le raccordement à un patient
peut être nocif pour le patient ou provoquer une ventilation inefficace. Lorsqu’un humidificateur
chauffant est employé, il convient de l’inclure dans le circuit en procédant à la vérification de
l’étanchéité.