Page 13-8 LTV
®
Series ventilator Operator’s Manual
p/n 10664, Rev. Y
3) To clean the exhalation valve and patient circuit, remove all gross particulate matter and bathe
for a minimum of 10 minutes in 50% water and 50% vinegar, KlenZyme, or another enzymatic
cleaner warmed to 95°F to 150°F (35°C to 65.5°C). Rinse gently for 2 minutes and use a low
flow air source to eliminate any residual fluid. Ultrasonic cleaning is not recommended. To clean
the water trap use a mild detergent solution followed by rinsing and drying with a low-flow air
source.
CAUTION
Cleaning the Water Trap – Do NOT use KlenZyme to clean Pulmonetic water traps. It will cause
deterioration of the water trap material.
ATTENTION
Nettoyage du piège à eau – NE PAS utiliser KlenZyme pour nettoyer les pièges à eau Pulmonetic.
Le cas échéant, le matériau composant le piège à eau se détériorera.
4) To high level disinfect the exhalation valve, patient circuit or water trap, remove all gross
particulate matter and bathe in a glutaraldehyde solution (e.g., Cidex (2%)) for 20 minutes. Rinse
gently for 2 minutes. Use a low flow air source to eliminate any residual fluid.
5) Exhalation valves, Patient Circuits and Water Traps are shipped clean, not sterile. Sterilization
of the exhalation valve, reusable patient circuit and water trap should follow individual institution
processes or guidelines.