Operator’s Manual LTV
®
Series ventilator Page 13-9
p/n 10664, Rev. Y
CAUTION
Reusable Patient Circuit Components - To avoid degradation of the reusable patient circuit
components, do not exceed the following constraints:
50 cleaning cycles or 1 year (whichever comes first)
Steam Autoclave:
Pressure: 20 PSIG
Temperature: 275°F (135°C)
Time: 6 minutes
Liquid Sterilizing Agent:
Do not use any of the following solutions to clean, disinfect, or sterilize the patient circuit:
Pasteurization:
30-minute warm water detergent and 30-minute 165°F (74°C) hot cycle.
Drying in a sterile drier for more than 1 hour or 140°F (59°C).
Gas (ETO):
Temperature: 131°F (55°C)
Care of the Exhalation Valve - The exhalation valve is delicate and may be damaged if;
Care is not exercised when handling or cleaning it.
Cleaning instruments or foreign bodies are inserted into it.
High-pressure gas nozzles are used to dry it.
Differential Pressure Ports - A low pressure air nozzle with flow less than 10 liters per minute
should be used for cleaning the differential pressure ports.
Patient Wye Installation – After cleaning, install the patient wye in the patient circuit so the proximal
sense lines are oriented up while operating.
Ketone Chlorinated solutions
Phenol (>5%) Chlorinated hydrocarbons
Inorganic acids Aromatic hydrocarbons
Formaldehyde Hypochlorite
Liquid agents containing more than
2% glutaraldehyde