Remote control commands
R&S
®
NRPxxS(N)
107User Manual 1177.5079.02 ─ 15
9.8.3 Excluding intervals
In the burst average and timeslot modes, you can define a time interval at the begin-
ning or at the end of an integration interval that is excluded from the measurement
result. See in Figure 9-6.
p
t
2
1
3
4
Figure 9-6: Excluding intervals
1 = Trigger level
2 = Excluding interval at the beginning
3 = Excluding interval at the end
4 = Integration interval
Remote commands:
[SENSe<Sensor>:]TIMing:EXCLude:STARt......................................................................107
[SENSe<Sensor>:]TIMing:EXCLude:STOP.......................................................................107
[SENSe<Sensor>:]TIMing:EXCLude:STARt <exclude_start>
Available in burst average and timeslot modes.
Sets a time interval at the beginning of bursts that is excluded from the measurement.
See Figure 9-6.
Parameters:
<exclude_start> Range: 0.0 to 1.0
*RST: 0.0
Default unit: Seconds
Manual operation: See "Start Exclude" on page 47
See "Start Exclude" on page 48
[SENSe<Sensor>:]TIMing:EXCLude:STOP <exclude_stop>
Available in burst average and timeslot modes.
Sets a time interval at the end of bursts that is excluded from the measurement. See
Figure 9-6.
Configuring basic measurement parameters