Remote control commands
R&S
®
NRPxxS(N)
129User Manual 1177.5079.02 ─ 15
Example:
SYST:ERR:CODE?
Query
0
Response: No errors have occurred since the error queue was
last read out.
Usage: Query only
SYSTem:ERRor:COUNt?
Queries the number of entries in the SCPI communication error queue.
Example:
SYST:ERR:COUN?
Query
1
Response: One error has occurred since the error queue was
last read out.
Usage: Query only
SYSTem:ERRor[:NEXT]?
Queries the SCPI communication error queue for the oldest entry and removes it from
the queue.
Returns an error number and a short description of the error.
Example:
SYST:ERR?
Query
0, 'no error'
Response: No errors have occurred since the error queue was
last read out.
Usage: Query only
SYSTem:FWUPdate <fwudata>
Loads new operating firmware into the power sensor. Rohde & Schwarz provides the
update file. For further details, see Chapter 7, "Firmware update", on page 57.
If you want to integrate a firmware update function in an application, see the example
given in Chapter 7.2.3, "Using remote control", on page 60.
Setting parameters:
<fwudata> <block_data>
Definite length arbitrary block data containing the direct copy of
the binary *.rsu file in the following format:
#
Single digit indicating how many digits follow to specify the size
of the binary file.
Number that specifies the size of the binary file.
Binary data
Configuring the system