EasyManua.ls Logo

R&S NRP Series - System:error:count; System:error[:Next]; System:fwupdate

R&S NRP Series
185 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
NRPxxS(N)
129User Manual 1177.5079.02 ─ 15
Example:
SYST:ERR:CODE?
Query
0
Response: No errors have occurred since the error queue was
last read out.
Usage: Query only
SYSTem:ERRor:COUNt?
Queries the number of entries in the SCPI communication error queue.
Example:
SYST:ERR:COUN?
Query
1
Response: One error has occurred since the error queue was
last read out.
Usage: Query only
SYSTem:ERRor[:NEXT]?
Queries the SCPI communication error queue for the oldest entry and removes it from
the queue.
Returns an error number and a short description of the error.
Example:
SYST:ERR?
Query
0, 'no error'
Response: No errors have occurred since the error queue was
last read out.
Usage: Query only
SYSTem:FWUPdate <fwudata>
Loads new operating firmware into the power sensor. Rohde & Schwarz provides the
update file. For further details, see Chapter 7, "Firmware update", on page 57.
If you want to integrate a firmware update function in an application, see the example
given in Chapter 7.2.3, "Using remote control", on page 60.
Setting parameters:
<fwudata> <block_data>
Definite length arbitrary block data containing the direct copy of
the binary *.rsu file in the following format:
#
Single digit indicating how many digits follow to specify the size
of the binary file.
Number that specifies the size of the binary file.
Binary data