SYSTem:INFO? [<item>]
Queries information about the power sensor.
If queried without parameters, the command returns all available information in the
form of a list of strings separated by commas.
If you want to query specific information, add the query parameter:
SYST:INFO? "<string>"
Query parameters:
<item> "Manufacturer", "Type", "Stock Number", "Serial", "SW Build",
"Sensor Name", "Hostname", "IP Address", "Technology", "Func-
tion", "MinPower", "MaxPower", "MinFreq", "MaxFreq", "Resolu-
tion", "Impedance", "Coupling", "Cal. Due Date", "Cal. Abs.",
"Cal. Refl.", "Cal. S-Para.", "Cal. S-Para. (User)", "Cal. Misc.",
"Cal. Temp.", "Cal. Lin.", "SPD Mnemonic", "TestLimit", "TestLi-
mit pd", "Uptime"
"Hostname" and "IP Address" are available only for LAN power
sensors.
Usage: Query only
Manual operation: See "Sensor Information" on page 51
SYSTem:INITialize
Sets the power sensor to the standard state, i.e. the default settings for all test parame-
ters are loaded in the same way as with *RST. The power sensor then outputs a com-
plete list of all supported commands and parameters. With the command, the remote-
control software can automatically adapt to the features of different types of sensors
with different functionality.
Usage:
Event
SYSTem:LANGuage <language>
Selects an emulation of a different command set.
Parameters:
<language> SCPI
*RST: SCPI
SYSTem:LED:COLor <color>
Effective if the operating mode of the LED is set to user-defined, SYSTem:LED:MODE
USER.
Sets the color and the flash code of the system status LED.
Parameters:
<color> Hexadecimal code described as
Configuring the system