Remote control commands
R&S
®
NRPxxS(N)
93User Manual 1177.5079.02 ─ 15
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:COUNt?
Available in continuous average mode.
Queries the number of results that are currently stored in the result buffer.
Example:
BUFF:COUN?
Usage: Query only
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:DATA?
Queries the results of the continuous average result buffer and returns them even if the
buffer is not full.
In contrast, FETCh<Sensor>[:SCALar][:POWer][:AVG]? returns a result only if
the buffer is full.
Usage:
Query only
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:SIZE <count>
Sets the size of the result buffer in continuous average mode.
You can enable the buffer using [SENSe<Sensor>:][POWer:][AVG:]BUFFer:
STATe.
Parameters:
<count> Range: 1 to 8192
*RST: 1
Example:
BUFF:SIZE 1
See Chapter 10.3, "Performing a buffered continuous average
measurement", on page 146.
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:STATe <state>
Enables or disables the buffered continuous average mode. If the buffer mode is
enabled, all results generated by trigger events are collected in the sensor until the buf-
fer is filled.
You can set the size of the buffer using [SENSe<Sensor>:][POWer:][AVG:
]BUFFer:SIZE.
Parameters:
<state> ON | OFF
*RST: OFF
Example:
BUFF:STAT OFF