Remote control commands
R&S
®
NRPxxS(N)
90User Manual 1177.5079.02 ─ 15
FORMat:SREGister <sregister>
Specifies which format is used for the return value of *STB?.
Parameters:
<sregister> ASCii | HEXadecimal | OCTal | BINary
*RST: ASCii
Example:
FORM:SREG ASC
FORMat[:DATA] [<data,length>, <length>]
Specifies how the R&S NRPxxS(N) sends the numeric data to the controlling host/
computer.
Parameters:
<data,length> <REAL,32 | 64>
REAL
Block of binary values, 32-bit or 64-bit each; so-called "SCPI
definite length block"
32 | 64
32-bit or 64-bit
If you omit the length, the R&S NRPxxS(N) sets the last used
length.
Example for REAL,32 format:
#6768000....<binary float values>....
Example for REAL,64 format:
#71536000....<binary float values>....
<data[,length]>
<ASCii[,0 to 12]>
ASCii
List of comma separated, readable values.
[,0 to 12]
Defines the number of decimal places.
The reset value 0 does not restrict the number of decimal pla-
ces.
Example for ASCii,4 format:
1.2938e-06, -4.7269e-11, ...
*RST: ASCii,0
9.7 Configuring the measurement modes
In the following, the settings needed for configuring a measurement mode are descri-
bed.
Further information:
●
Chapter 9.8, "Configuring basic measurement parameters", on page 102
●
Chapter 9.5, "Controlling the measurement", on page 73
Configuring the measurement modes