Remote control commands
R&S
®
NRPxxS(N)
135User Manual 1177.5079.02 ─ 15
Example:
SYST:SERR:LIST?
Query
0,"reported at uptime:2942; notice;
auto-averaging exceeded maximum time;
Notification"
Response
Usage: Query only
SYSTem:SUTime <update_time>
Effective only in the NRP legacy mode. Relevant only in continuous measurement
mode, INITiate:CONTinuousON.
Sets the status update time. That is the maximum rate in which the power sensor can
output measurement results.
Parameters:
<update_time> Range: 0.0 to 10.0
*RST: 10e-3
Default unit: Seconds
SYSTem:TLEVels?
Queries the possible power test levels of the power sensor.
Usage:
Query only
SYSTem:TRANsaction:BEGin
Starts a series of settings.
Usage:
Event
SYSTem:TRANsaction:END
Ends a series of settings.
Usage:
Event
SYSTem[:SENSor]:NAME <sensorname>
Sets the name of the power sensor according to your requirements. The specified
name is displayed in the web user interface of the network power sensors.
The name that you specify here is independent from the hostname of the power sen-
sor. However, if you do not specify a name, the hostname is used as default.
Configuring the system