R&S NRP Series - System:led:mode; System:minpower; System:parameters

R&S NRP Series
185 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
NRPxxS(N)
132User Manual 1177.5079.02 ─ 15
0x0krrggbb
with
k = 0: steady on; k = 1: slowly flashing; k = 2: fast flashing
rr = red
gg = green
bb = blue
In NRP legacy communication, the parameter is a standard deci-
mal number, representing the corresponding hexadecimal code.
Range: 0x00 to 0x02FFFFFF
*RST: 0x00A0A0A0
Example:
SYST:LED:MODE USER
Sets the system status LED operating mode to user.
SYST:LED:COL #H01a00000
The LED flashes slowly in red.
SYSTem:LED:MODE SENSor
Sets the system status LED operating mode back to the sensor
internal settings.
SYSTem:LED:MODE <mode>
Sets whether the color of the system status LED is controlled by the power sensor firm-
ware or by the user settings.
For more information, see SYSTem:LED:COLor.
Parameters:
<mode> USER | SENSor
*RST: SENSor
SYSTem:MINPower?
Queries the lower power measurement limit.
This value changes if [SENSe<Sensor>:]CORRection:SPDevice:STATe is set to
ON. The lower measurement limit refers to the sensor or to the combination of a sensor
and the components connected ahead of it.
Use this query to determine a useful resolution for the result display near the lower
measurement limit.
Usage:
Query only
SYSTem:PARameters?
Returns an XML-output containing all commands with the following information, if avail-
able for the command:
Default value
Configuring the system