Remote control commands
R&S
®
NRPxxS(N)
66User Manual 1177.5079.02 ─ 15
Example:
Definition: [SENSe<Sensor>:]AVERage:COUNt:AUTO:NSRatio <nsr>
Command: AVER:COUN:AUTO:NSR 0.01
Special characters | and { }
|
A vertical bar in parameter definitions indicates alternative possibilities in the sense of
"or". The effect of the command differs, depending on which parameter is used.
Example:
Definition: INITiate:CONTinuous ON | OFF
Command INITiate:CONTinuous ON starts the measurements
Command INITiate:CONTinuous OFF stops the measurements
{ } Parameters in braces can be included in the command once, several times or not at all.
9.3 Common commands
The common commands are taken from the IEEE 488.2 (IEC 625–2) standard. The
headers of these commands consist of an asterisk * followed by three letters.
*CLS...............................................................................................................................66
*ESE...............................................................................................................................67
*ESR?.............................................................................................................................67
*IDN?..............................................................................................................................67
*IST?.............................................................................................................................. 67
*OPC..............................................................................................................................67
*OPT?.............................................................................................................................68
*PRE.............................................................................................................................. 68
*RCL...............................................................................................................................68
*RST...............................................................................................................................68
*SAV...............................................................................................................................68
*SRE.............................................................................................................................. 69
*STB?.............................................................................................................................69
*TRG.............................................................................................................................. 69
*TST?.............................................................................................................................69
*WAI............................................................................................................................... 69
*CLS
Clear status
Resets the following:
●
Status byte (STB)
●
Standard event register (ESR)
●
EVENt part of the QUEStionable and the OPERation register
●
Error/event queue
Common commands