Browser-based user interface
R&S
®
NRPxxS(N)
53User Manual 1177.5079.02 ─ 15
Remote command:
TRIGger:SLOPe
Trigger Level
Sets the trigger threshold for internal triggering derived from the test signal.
Remote command:
TRIGger:LEVel on page 86
TRIGger:LEVel:UNIT on page 87
Trigger Delay
Sets the delay between the trigger event and the beginning of the actual measurement.
Remote command:
TRIGger:DELay
Dropout
With a positive (negative) trigger slope, the dropout time is the minimum time for which
the signal must be below (above) the power level defined by "Trigger Level". See also
Chapter 9.5.2.4, "Dropout time", on page 76.
Remote command:
TRIGger:DTIMe
Holdoff
Sets the hold-off time, a period after a trigger event during which all trigger events are
ignored. See also Chapter 9.5.2.5, "Hold-off time", on page 77.
Remote command:
TRIGger:HOLDoff
Hysteresis
Sets the hysteresis in dB. A trigger event occurs, if the trigger level:
●
Falls below the set value on a rising slope.
●
Rises above the set value on a falling slope.
Thus, you can use this setting to eliminate the effects of noise in the signal for the edge
detector of the trigger system.
Remote command:
TRIGger:HYSTeresis
6.5.4 System settings
Describes the parameters of the general network environment and specific identifica-
tion parameters of the power sensor in the network.
Further information:
●
Chapter 9.12, "Configuring the system", on page 127
Access: main dialog of the web user interface > navigation pane > "System"
Settings