Remote control commands
R&S
®
NRPxxS(N)
133User Manual 1177.5079.02 ─ 15
●
Minimum value
●
Maximum value
●
Parameters
●
Limits
Each command is shortened to a command token, consisting only of the mnemonics
short form. For example, CALibration:DATA is shortened to CALDATA as command
token.
Usage: Query only
SYSTem:PARameters:DELTa?
Returns an XML-output containing all commands that differ from the defined default
status set by *RST on page 68.
The commands are accompanied by the same information as for SYSTem:
PARameters?.
Usage:
Query only
SYSTem:PRESet
Resets the power sensor.
The command essentially corresponds to the *RST command, with the exception that
the settings of the following commands are persistently held:
INITiate:CONTinuous
SENSe:AVERage:TCONtrol
SENSe:TRACe:AVERage:TCONtrol
Usage:
Event
SYSTem:REBoot
Reboots the power sensor.
Usage:
Event
SYSTem:RESTart
Restarts the firmware of the power sensor.
Usage:
Event
SYSTem:RUTime <update_time>
Effective only in the NRP legacy mode. Relevant only in continuous measurement
mode, INITiate:CONTinuous ON.
Configuring the system