Remote control commands
R&S
®
NRPxxS(N)
102User Manual 1177.5079.02 ─ 15
Manual operation: See "Trace Points" on page 49
[SENSe<Sensor>:]TRACe:REALtime <state>
Available in trace mode.
If disabled, each measurement from the power sensor is averaged. If enabled, only
one sampling sequence per measurement is recorded, thus increasing the measure-
ment speed. With a higher measurement speed, the measured values of an individual
measurement are immediately delivered.
The averaging filter is not used, so the following settings are ignored:
●
[SENSe<Sensor>:]TRACe:AVERage[:STATe]
●
[SENSe<Sensor>:]TRACe:AVERage:COUNt
Parameters:
<state> *RST: OFF
[SENSe<Sensor>:]TRACe:TIME <time>
Sets the trace length, time to be covered by the trace sequence. This time period is
divided into several equal intervals, in which the average power is determined. The
number of intervals equals the number of trace points, which is set using
[SENSe<Sensor>:]TRACe:POINts.
Parameters:
<time> Range: 10.0e-6 to 3.0
*RST: 0.01
Default unit: Seconds
Manual operation: See "Trace Time" on page 49
9.8 Configuring basic measurement parameters
The following section describes the settings common for several measurement modes.
9.8.1 Configuring auto averaging
Describes the commands for automatic averaging in continuous average and burst
average measurements.
Remote commands:
[SENSe<Sensor>:]AVERage:COUNt................................................................................103
[SENSe<Sensor>:]AVERage:COUNt:AUTO......................................................................103
[SENSe<Sensor>:]AVERage:COUNt:AUTO:MTIMe...........................................................103
[SENSe<Sensor>:]AVERage:COUNt:AUTO:NSRatio.........................................................104
[SENSe<Sensor>:]AVERage:COUNt:AUTO:RESolution.....................................................104
[SENSe<Sensor>:]AVERage:COUNt:AUTO:SLOT............................................................ 104
Configuring basic measurement parameters